IEC 60751 · 1000 Ω · Long-Run Cable · AS9100D

PT1000
RTD Sensor

PT1000 platinum resistance thermometers with R₀ = 1000.000 Ω at 0°C per IEC 60751:2022. Identical α coefficient and tolerance classes as PT100 but 10× higher base resistance — making lead resistance effects 10× less significant.

Technical Data

Specification Summary
Nominal Resistance1000.000 Ω at 0°C (ITS-90)
α Coefficient3.85055×10⁻³ °C⁻¹ (IEC 60751)
Temperature Range–200°C to +850°C
Class AA Tolerance±(0.100 + 0.0017|t|)°C
Class A Tolerance±(0.150 + 0.0020|t|)°C
Class B Tolerance±(0.300 + 0.0050|t|)°C
Resistance at +100°C1385.06 Ω
Resistance at –100°C602.56 Ω
Lead Resistance Effect0.1 Ω lead wire error = 0.026°C at PT1000 vs 0.26°C at PT100
Wire Configurations2-wire, 3-wire, 4-wire
Sheath Materials316 SS, Inconel 600, PTFE
OD Range3.18 mm to 12.7 mm standard
DocumentationAS9100D COC + material certs standard; NIST cal on request
Technical Notes
PT1000 uses the identical IEC 60751 Callendar-Van Dusen equation and tolerance classes as PT100 — only the nominal resistance (R₀) differs. The 10× higher resistance means each ohm of lead wire resistance produces only 0.026°C of error vs 0.26°C for PT100. This makes PT1000 the preferred choice for 2-wire installations with cable runs over 10–15 m, battery-powered dataloggers, and applications where instruments lack 3 or 4-wire RTD input capability.

Common Applications

Building automation and HVAC with long cable runs (2-wire)
Battery-powered environmental monitoring and dataloggers
Field-installed sensors where conduit is limited to 2-wire
Remote process temperature monitoring
Food cold chain and refrigeration with extended cable lengths