Nominal Resistance1000.000 Ω at 0°C (ITS-90)
α Coefficient3.85055×10⁻³ °C⁻¹ (IEC 60751)
Temperature Range–200°C to +850°C
Class AA Tolerance±(0.100 + 0.0017|t|)°C
Class A Tolerance±(0.150 + 0.0020|t|)°C
Class B Tolerance±(0.300 + 0.0050|t|)°C
Resistance at +100°C1385.06 Ω
Resistance at –100°C602.56 Ω
Lead Resistance Effect0.1 Ω lead wire error = 0.026°C at PT1000 vs 0.26°C at PT100
Wire Configurations2-wire, 3-wire, 4-wire
Sheath Materials316 SS, Inconel 600, PTFE
OD Range3.18 mm to 12.7 mm standard
DocumentationAS9100D COC + material certs standard; NIST cal on request
Technical Notes
PT1000 uses the identical IEC 60751 Callendar-Van Dusen equation and tolerance classes as PT100 — only the nominal resistance (R₀) differs. The 10× higher resistance means each ohm of lead wire resistance produces only 0.026°C of error vs 0.26°C for PT100. This makes PT1000 the preferred choice for 2-wire installations with cable runs over 10–15 m, battery-powered dataloggers, and applications where instruments lack 3 or 4-wire RTD input capability.