Nominal Resistance100.000 Ω at 0°C (ITS-90)
α Coefficient3.85055×10⁻³ °C⁻¹ (IEC 60751)
Temperature Range–200°C to +850°C (sheathed)
Class AA Tolerance±(0.100 + 0.0017|t|)°C — –50 to +250°C only
Class A Tolerance±(0.150 + 0.0020|t|)°C — –100 to +450°C
Class B Tolerance±(0.300 + 0.0050|t|)°C — –196 to +600°C
Class C Tolerance±(0.600 + 0.010|t|)°C — –196 to +600°C
Resistance at +100°C138.506 Ω (Callendar-Van Dusen)
Resistance at –100°C60.256 Ω
Resistance at +400°C247.090 Ω
Wire Configurations2-wire, 3-wire, 4-wire
Sheath Materials304 SS, 316 SS, Inconel 600, Hastelloy C-276
OD Range3.18 mm (1/8") to 12.7 mm (1/2") standard
DocumentationAS9100D COC + material certs standard; NIST cal on request
Technical Notes
PT100 (100 Ω at 0°C) is the most common RTD value worldwide. IEC 60751:2022 governs the resistance-temperature relationship using the Callendar-Van Dusen equation: R(T) = R₀[1+AT+BT²] for T ≥ 0°C, with A = 3.9083×10⁻³, B = –5.775×10⁻⁷. 4-wire measurement is recommended for the highest accuracy — it eliminates lead resistance error entirely. 3-wire is acceptable for runs under ~30 m when lead resistance is balanced.